Test method for liquid leakage at syringe piston, ISO 7886-1 Annex D

The standard ISO 7886-1 : 1993 Annex D specifies a test method for liquid leakage at syringe piston under compression. The procedure includes the following steps: 1. Draw into the syringe a volume of water exceeding the nominal capacity of the syringe. 2. Expel air and adjust the volume of water in the syringe to the nominal capacity. 3. Seal the syringe nozzle using the cap provided. 4. Apply and axial force and maintain it for 30 seconds. 5. Examine the syringe for leakage of water past the piston seals.

For this particular test, a force value 20% below the level when a failure of the cap was registered was the force selected to constantly apply during 30 seconds. This period started after the selected force level was reached for the first time. After this test, a leakage at the nozzle cap should be visually detected, a leakage either partial or total can be also monitored if a drop in the force level is noticed during the test. The Instron Syringe fixture CP102517 provides an ideal solution for a liquid leakage test, the Instron® software Bluehill® also provides the required control tools to easily perform a force hold during this test.


이 솔루션 정보

관련 표준: ISO 7886-1
시편 종류: Syringe
재료: 바이오메디컬
시험 유형: 압축시험
비즈니스 부문: 바이오메디컬/의료산업/헬스케어
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
이 솔루션에 관심이 있으십니까?
Contact Us

관련 문헌

Bluehill 3 Testing Software Brochure   Bluehill 3 Testing Software Brochure
Syringe Test Fixture   Syringe Test Fixture
Syringe Test Fixture for ISO-7886-1   Syringe Test Fixture for ISO-7886-1

인스트론 연구소

Instron 애플리케이션 연구소에서는 R&D를 비롯하여 고객이 적합한 장비를 선택할 수 있도록 고객의 샘플 시험 작업이 이루어집니다.
지금 바로 애플리케이션 전문가와 상의하여 귀하께 가장 적합한 솔루션을 찾아 보십시오! 어플리케이션 전문가와 상담하십시오.