Fatigue Testing of Pre-Angled Endosseous Dental Implants in a Fluid Bath (ISO 14801)

Pre-Angled dental implants are designed to help dental surgeons make the best use of the existing bone or to remove the need for bone grafting to occur, while providing a secure restorative platform.

The latest revision of the ISO 14801 standard for fatigue testing specifies a method for testing of these pre-angled dental implants in a laboratory situation. The standard requires the implant to be aligned so that the angle of loading is 10° greater than between the implant axis and the axis of the angled portion of the connecting part.

With an array of implant sizes and geometries, angled dental implants can be mounted into a variable angle fixture for testing to the ISO 14801 standard. The fixture needs to ensure that the mounting does not over constraint the implant nor generate large lateral forces that can damage the load cell or the testing machine. Pre-angled dental implants should be tested in ambient air using a frequency of 15 Hz.

As with straight dental implants, pre-angled dental implants can also be tested in an aqueous environment, if corrosion fatigue is expected to occur. The fixture can be mounted within a fluid bath to test the implant in saline solution or another physiological medium and the temperature maintained at 37 °C.

Our E1000, E3000, or ElectroPuls™ linear motor test instruments provide an ideal platform for testing and characterizing dental implant performance. Alternatively, the 8870 series of servohydraulic fatigue testing systems can be used. If studies of torque required for dental implants need to be undertaken, we suggest using the ElectroPuls E10000 Linear-Torsion test instrument or our 8874 servohydraulic axial-torsion testing system to conduct compression and torsion testing.

We recommend that you review the standard to fully understand its requirements.

Interested in other mechanical tests for dental materials or dental implants? We have a variety of different testing equipment solutions for dentistry.


이 솔루션 정보

관련 표준: ISO 14801
시편 종류: 임플란트
재료: 바이오메디컬 | 생체 재료
시험 유형: 고주기피로시험
비즈니스 부문: 학술분야 | 바이오메디컬/의료산업/헬스케어 | 시험 수탁 기관
Dental Implant in Bio Bath
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
이 솔루션에 관심이 있으십니까?
Contact Us

관련 문헌

Solutions for Biomedical Testing   Solutions for Biomedical Testing

인스트론 연구소

Instron 애플리케이션 연구소에서는 R&D를 비롯하여 고객이 적합한 장비를 선택할 수 있도록 고객의 샘플 시험 작업이 이루어집니다.
지금 바로 애플리케이션 전문가와 상의하여 귀하께 가장 적합한 솔루션을 찾아 보십시오! 어플리케이션 전문가와 상담하십시오.